Miniaturized High-Resolution Time-of-Flight Mass Spectrometer MULTUM-S II with an Infinite Flight Path

作者:Shimma Shuichi*; Nagao Hirofumi; Aoki Jun; Takahashi Keiji; Miki Shinichi; Toyoda Michisato
来源:Analytical Chemistry, 2010, 82(20): 8456-8463.
DOI:10.1021/ac1010348

摘要

A new miniature multiturn time-of-flight (TOF) analyzer "MULTUM-S II" has been designed and constructed. This instrument consists of an electron ionization source, the multiturn TOF ion optics, a detector, vacuum system, and electronic circuits. The multiturn TOF analyzer consists of four electrostatic toroidal sectors and two additional electric toroidal sectors for the purpose of ion injection/ejection. The size and weight of the system is less than 50 cm x 57 cm x 30 cm and 35 kg (including vacuum pumps and electronic circuits). The multiturn TOF analyzer is capable of high mass resolution because of its infinite flight path utilizing perfect space and time focused closed flight orbit. To evaluate the resolution in MULTUM-S II, separation of pyridine ((12)C(5)H(5)N) and the isotopic component of benzene ((13)C(12)C(5)H(6)) was performed at a mass resolution of about 20 000. Another performance characteristic of the MULTUM-S II was demonstrated by the separation of the greenhouse gas doublet CO(2) and N(2)O (Delta m = 0.0113 Da). While the mass difference is a mere 0.01 Da, the instrument could easily separate the two peaks at a calculated mass resolution of 31 600. The MULTUM-S II offers high mass resolution mass spectrometry in a miniaturized/portable enclosure.

  • 出版日期2010-10-15