A Novel Technique for Monitoring of Fast Variations in Brain Oxygen Tension Using an Uncoated Fluorescence Quenching Probe (Foxy AL-300)

作者:Klein Klaus Ulrich*; Boehme Stefan; Hartmann Erik Kristopher; Szczyrba Marc; David Matthias; Markstaller Klaus; Engelhard Kristin
来源:Journal of Neurosurgical Anesthesiology, 2011, 23(4): 341-346.
DOI:10.1097/ANA.0b013e31822cf893

摘要

Background: A novel uncoated fluorescence quenching probe allows fast measurement of oxygen tension in vessels and tissue. The present study reports the first use of the technology for dual measurements of arterial (paO(2)) and brain tissue oxygen tension (ptiO(2)) during hypoxic challenge in a pig model.
Methods: Eight pigs were anesthetized using fentanyl and propofol. Fluorescence quenching pO(2) probes (Foxy AL-300, Ocean Optics, Dunedin, FL) were placed in the ascending aorta (Foxy-paO(2)) and subcortically at 14 mm in brain tissue (Foxy-ptiO(2)). As reference, a clark-type electrode probe (Licox-ptiO(2)) was placed into brain tissue close to the Foxy probe (Licox, Integra Neurosciences, Plainsboro, NJ). Measurements were taken at baseline (FiO(2) 1.0), during episodes of apnea, and during recovery (FiO(2) 1.0). Statistics: descriptive results.
Results: Individual Foxy-paO(2), Foxy-ptiO(2), and Licox-ptiO(2) courses were related to episodes of apnea. The response time of the Foxy measurements was 10Hz. Baseline values at FiO(2) 1.0 were Foxy-paO(2) 520 +/- 120 mm Hg, Foxy-ptiO(2) 62 +/- 24 mm Hg, and Licox-ptiO(2) 55 +/- 29 mm Hg; apnea values were Foxy-paO(2) 64 +/- 10 mm Hg, Foxy-ptiO(2) 37 +/- 12 mm Hg, and Licox-ptiO(2) 31 +/- 16 mm Hg; recovery values at FiO(2) 1.0 were Foxy-paO(2) 478 +/- 98 mm Hg, Foxy-ptiO(2) 78 +/- 26 mm Hg, and Licox-ptiO(2) 62 +/- 32 mm Hg.
Conclusions: The present study demonstrates the feasibility of pO(2) measurements in macrocirculation and cerebral microcirculation using a novel uncoated fluorescence quenching probe. The technology allows for real-time investigation of pO(2) changes at a temporal resolution of 0.05 to 10 Hz.

  • 出版日期2011-10