Mapping the strain distribution on the proximal femur with titanium and flexible-stemmed implants using digital image correlation

作者:Tayton E*; Evans S; O'Doherty D
来源:Journal of Bone and Joint Surgery-British Volume, 2010, 92B(8): 1176-1181.
DOI:10.1302/0301-620X.92B8.23553

摘要

We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants. This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.