摘要
This study evaluated spatial Phi(th) inside a 70 kg water phantom using the NAA method. Fifty indium foils were placed inside the water phantom and exposed under 15 MV LINAC for 2.5 min to yield the 10 Gy X-ray dose. The Phi(th) value at the isocenter of the water was 1.03 x 10(6) n cm(-2)/Gy-X, and the maximum quantity of Phi(th) appeared at the water surface along the z-axis, 3.99 x 10(6) n cm(-2)/Gy-X. The thermal neutron dose at isocenter of the water phantom occupied approximately 0.151% of the whole photo and neutron dose.