摘要

Sr-90 is a widely determined radionuclide for environmental purposes, nuclear waste control, and can be also monitored in coolants in nuclear reactor plants. In the developed method, the ICP-MS detection was employed together with sample processing in sequential injection analysis (SIA) setup, equipped with a lab-on-valve with mechanized renewal of sorbent bed for solid-phase extraction. The optimized conditions of determination included preconcentration of Sr-90 on cation-exchange column and removal of different type of interferences using extraction Sr-resin. The limit of detection of the developed procedure depends essentially on the configuration of the employed ICP-MS spectrometer and on the available volume of the sample to be analyzed. For 1 L initial sample volume, the method detection limit (MDL) value was evaluated as 2.9 ppq (14.5 Bq L-1). The developed method was applied to analyze spiked river water samples, water reference materials, and also simulated and real samples of the nuclear reactor coolant.

  • 出版日期2017-7-1