Degree of polarization uniformity with high noise immunity using polarization-sensitive optical coherence tomography

作者:Makita Shuichi; Hong Young Joo; Miura Masahiro; Yasuno Yoshiaki*
来源:Optics Letters, 2014, 39(24): 6783-6786.
DOI:10.1364/OL.39.006783

摘要

A new metric representing polarization uniformity is presented. Noise corrected degree of polarization uniformity (DOPU) is computed from polarization-sensitive optical coherence tomography (OCT), and selectively visualizes tissue with the multiple scattering, such as highly pigmented tissues. The new metric is designed to be minimally sensitive to systematic additive noise. The performance of this new metric is analyzed by numerical simulation and in vivo human retinal imaging, using Jones matrix OCT. The new metric exhibited only a small dependency on the signal-to-noise ratio. Selective in vivo visualization of pigmented tissues in the human retina is demonstrated, with cross sectional and en-face images.

  • 出版日期2014-12-15