摘要

Modern full-field digital mammography (FFDM) units display the mean glandular dose (MGD) and the entrance or incident air kerma (K) to the breast following each exposure. Information on how these values are calculated is limited and knowing how displayed MGD values compare and correlate to conventional Monte-Carlo-based methods is useful. From measurements done on polymethyl methacrylate (PMMA) phantoms, it has been shown that displayed and calculated MGD values are similar for thin to medium thicknesses and appear to differ with larger PMMA thicknesses. As a result, a multiple linear regression analysis on the data was performed to generate models by which displayed MGD values on the two FFDM units included in the study may be converted to the Monte-Carlo values calculated by conventional methods. These models should be a useful tool for medical physicists requiring MGD data from FFDM units included in this paper and should reduce the survey time spent on dose calculations.

  • 出版日期2013-4