A Monte Carlo dosimetry study using Henschke applicator for cervical brachytherapy

作者:Yu Pei Chieh; Chao Tsi Chian; Lee Chung Chi; Wu Ching Jung; Tung Chuan Jong*
来源:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment , 2010, 619(1-3): 411-414.
DOI:10.1016/j.nima.2010.01.079

摘要

In recent years the Henschke applicator has been widely used for gynecologic patients treated by brachytherapy in Taiwan. However, the commercial brachytherapy planning system did not properly evaluate the dose perturbation caused by the Henschke applicator. Since the European Society for Therapeutic Radiology and Oncology advised that the effect of source shielding should be incorporated into the brachytherapy planning system, it required calculation and comparison of the dose distribution around the applicator. This study used the Monte Carlo MCNP code to simulate the dose distribution in a water phantom that contained the Henschke applicator with one tandem and two ovoids. Three dwell positions of a high dose rate (192)Ir source were simulated by including and excluding the applicator. The mesh tally option of the MCNP was applied to facilitate the calculation of a large number of tallies in the phantom. The voxel size effect and the charge particle equilibrium were studied by comparing the results calculated with different tally options. The calculated results showed that the brachytherapy planning system overestimated the rectal dose and that the shielding material in the applicator contributed more than 40% to the rectal dose.