MRC-SPECT: A sub-500 mu m resolution MR-compatible SPECT system for simultaneous dual-modality study of small animals

作者:Cai Liang*; Lai Xiaochun; Shen Zengming; Chen Chin Tu; Meng Ling Jian
来源:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment , 2014, 734: 147-151.
DOI:10.1016/j.nima.2013.03.080

摘要

In this paper, we will report the development of an ultrahigh resolution MR-compatible SPECT system that can be operated inside a pre-existing clinical MR scanner for simultaneous dual-modality imaging of small animals. This system is constructed with 40 small-pixel CdTe detector modules assembled in a fully stationary ring SPEC geometry. We have experimentally demonstrated that this system is capable of providing an imaging resolution of <500 mu m when operating inside MR scanners. We will report the design, construction of the MRI-compatible SPECT system, including the detector technology, collimator, system development and so on The first imaging results obtained with this newly constructed SPECT system will also be reported.

  • 出版日期2014-1-11