In this study, the performance of linogram acquisition was investigated for the reconstruction of images from undersampled data using parallel imaging methods. The point spread PSF) of linogram sampling was analyzed for image sharpness and artifacts. Generalized auto-calibrating partially parallel acquisition was implemented for this new sampling scheme, and images were reconstructed with high acceleration rates. The results were compared with conventional radial sampling methods using simulations and phantom experiments at 3T. Additionally, a human volunteer was scanned at 7T. The results demonstrated that the PSF was sharper and the mean artifact power was lower for linogram sampling compared with radial sampling. Results of simulations and phantom experiments were in accord with the findings of the PSF analysis. In simulations, errors in the reconstructed images were lower for linogram sampling. In phantom experiments, fine details and sharp edges were preserved for linogram sampling, while details were blurred for radial sampling. The in vivo human study demonstrated that linogram sampling could provide high quality images of anatomy, even at high acceleration rates. Linogram sampling not only possesses the advantages of radial sampling, such as reduced sensitivity to motion and higher acceleration rates, but it also provides sharper images with fewer artifacts. Moreover, it is less prone to off-resonance artifacts compared with radial sampling.