A method to stabilize the temperature dependent performance of G-APD arrays

作者:Huh Yoonsuk; Choi Yong*; Jung Jin Ho; Jung Jiwoong
来源:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment , 2015, 772: 83-88.
DOI:10.1016/j.nima.2014.11.006

摘要

This paper presents a compensation method to stabilize the temperature dependent performance of Geiger-mode Avalanche Photodiode (G-APD) arrays for Positron Emission Tomography (PET). The compensation method is used to identify the bias voltage range that provides stable performance even at different temperatures using the G-APD's characteristics, and to control the photo-peak variation as a Function of temperature using the preamplifier gain within the identified bias voltage range. A pair of GAPD detectors and temperature sensors were located in the temperature chamber and the preamplifiers which can control the gain of the detectors using the digital potentiometer were positioned outside the chamber. The performance of the G-APD detector, especially energy resolution and coincidence timing resolution, was characterized as a function of bias voltage at different temperatures from 20 degrees C to 40 degrees C at 5 degrees C increments; the energy resolution, coincidence Liming resolution, and photo-peak position of all channels of G-APD PET defectors before and after the preamplifier gain correction were then measured and compared. The results of this study demonstrated that the optimal bias voltage range providing the good energy and coincidence Liming resolution, 12.1 +/- 1.2% and 1.30 +/- 0.09 ns, respectively, could be identified at the temperature range and the photo-peak variation and the performance at different temperatures could be stabilized by adjusting the preamplifier gain within the identified bias voltage range. We concluded the proposed method to be reliable and useful for the development of the PET system using G-APD arrays.

  • 出版日期2015-2-1