摘要

Stomach contractions are initiated and coordinated by electrical events termed slow waves, and slow wave abnormalities contribute to gastric motility disorders. Recently, flexible printed circuit board (PCB) multi-electrode arrays were introduced, facilitating high-resolution mapping of slow wave activity in humans. However PCBs with gold contacts have shown a moderately inferior signal quality to previous custom-built silver-wire platforms, potentially limiting analyses. This study determined if using silver instead of gold contacts improved flexible PCB performance. In a salt-bath test, modestly higher stimulus amplitudes were recorded from silver PCBs (mean 312, s.d. 89 mu V) than those from gold (mean 281, s.d. 85 mu V) (p < 0.001); however, the signal-to-noise ratio (SNR) was similar (p = 0.26). In eight in vivo experimental studies, involving gastric serosal recordings from five pigs, no silver versus gold differences were found in terms of slow wave amplitudes (mean 677 versus 682 mu V; p = 0.91), SNR (mean 8.8 versus 8.8 dB; p = 0.94) or baseline drift (NRMS; mean 12.0 versus 12.1; p = 0.97). Under the prescribed conditions, flexible PCBs with silver or gold contacts provide comparable results in vivo, and contact material difference does not explain the performance difference between current-generation slow wave mapping platforms. Alternative explanations for this difference and the implications for electrode design are discussed.

  • 出版日期2011-3