Application of monolithic capillary half focusing X-ray lens in micro-EXAFS analytical technology

作者:Sun Tian Xi*; Liu Zhi Guo; Wei Shi Qiang; He Bo; Wei Zheng; Yang Jun; Zeng Yi; Lin Xiao Yan; Chu Xue Lian; Ding Xun Liang
来源:Chinese Physics C, 2007, 31(10): 967-971.

摘要

A monolithic capillary half focusing X-ray lens (MCHFXRL) is used to focus the synchrotron radiation. The magnitudes of the focal spot size and gain of the synchrotron radiation microbeam focused by the MCHFXRL are in the orders of 10 mu m and 10(3), respectively. In the energy range from 5.5 to 11.5keV, the focal spot size varies from 88 to 29 mu m; the transmission efficiency varies from 26.1% to 20.5%; the translation of the vertical position of the focal spot is 3 mu m; the change of the output focal distance is 155 mu m. On the basis of the investigation of the performances of the MCHFXRL, the application of the synchrotron radiation microbeam in micro-extended-X-ray-absorption-fine-structure (micro-EXAFS) analysis is studied.