AN INTERCOMPARISON FOR NIRS AND NYU PASSIVE THORON GAS DETECTORS AT NYU

作者:Sorimachi Atsuyuki; Ishikawa Tetsuo; Tokonami Shinji*; Chitta**** Passa****; Harley Naomi H
来源:Health Physics, 2012, 102(4): 419-424.
DOI:10.1097/HP.0b013e31823ecd5c

摘要

An intercomparison on thoron (Rn-220) measurement was carried out between National Institute of Radiological Sciences, Japan (NIRS), and New York University School of Medicine, USA (NYU). The measurements of Rn-220 concentration at NIRS and NYU were performed by using the scintillation cell method and the two-filter method, respectively, as the standard measurement method. Three types of alpha track detectors based on passive radon (Rn-222)-Rn-220 discriminative measurement technique were used: Raduet and Radopot detectors were used at NIRS, and four-leaf detectors were used at NYU. In this study, the authors evaluated Rn-220 concentration variation in terms of run for exposure, measurement method, and exposure chamber. The detectors were exposed to Rn-220 gas with approximately 15 kBq m(-3) during the period from 0.75 to 3 d. As a result, the variation of each measurement method among these exposure runs was comparable to or less than that for the two-filter method. Agreement between the standard measurement methods of NIRS and NYU was observed to be about 10%, as is the case with the passive detectors. The Raduet detector showed a large variation in the detection response between the NIRS and NYU chambers, which could be related to different traceability. Health Phys. 102(4):419-424; 2012

  • 出版日期2012-4