A new mammography dosimetric phantom

作者:Almeida C D*; Coutinho C M C; Dantas B M; Peixoto J E; Koch H A
来源:Radiation Protection Dosimetry, 2012, 151(1): 196-198.
DOI:10.1093/rpd/ncr457

摘要

Breast phantoms produced with tissue-equivalent materials are used in an attempt to simulate glandular and adipose tissues, in terms of X-ray attenuation and density. In this work, a set of breast tissue-equivalent phantoms (BTE phantoms) with semicircular shapes of different thicknesses and compositions were produced. Such phantoms may be used in the measurement of the incident air kerma (K-i) and the mean glandular dose (D-G) delivered to patients undergoing mammography. To characterise the materials used to produce the phantoms, a series of 17-keV X-ray attenuation coefficient measurements were performed. The carbonnitrogenhydrogen elemental composition and the densities of the tissue-equivalent materials were also determined and compared with values available in the literature. Linear attenuation coefficients of 0.724 and 0.923 cm(1) were determined, respectively, for adipose and glandular tissues. Such values agree with data available in the literature. On the basis of the results obtained in this work, it is suggested that BTE phantoms are used instead of polymethyl methacrylate phantoms to select exposure parameters (kV, mAs and target/filter combination) specific for breast glandularities from 0 to 50 in the optimisation of doses in mammography.

  • 出版日期2012-8