Applications of MOSkin dosimeters for quality assurance in gynecological HDR brachytherapy: An in-phantom feasibility study

作者:Romanyukha Anna A.*; Carrara Mauro; Tenconi Chiara; Mazzeo Davide; Rossi Giulio; Borroni Marta; Pignoli Emanuele; Cerrotta Annamaria; Cutajar Dean; Petasecca Marco; Lerch Michael; Bucci Joseph; Rosenfeld Anatoly B.
来源:Radiation Measurements, 2017, 106: 399-404.
DOI:10.1016/j.radmeas.2016.12.010

摘要

In vivo dosimetry (IVD) is an excellent mode of treatment verification and detection of possible over-exposures. A feasibility study was conducted to evaluate a proposed IVD procedure for gynecological HDR brachytherapy procedure quality assurance. MOSkin dosimeters were selected due to their small size and capability of measuring steep dose gradients, such as those relevant in HDR brachytherapy procedures. Two in-phantom experiments measuring dose with MOSkins at the position simulating the rectal wall were conducted employing a cylindrical single channel applicator and a multichannel applicator. Three MOSkins were incorporated onto a rectal catheter, which was then attached to the applicator, separated by a small wooden plaque, simulating the vaginal to rectal wall distance and fixing catheter position. This setup was inserted into a water phantom and three treatment plans prescribing 300 cGy to 3 different targets were assigned with various dose distributions. Each treatment was administered three times, and doses measured by the MOSkins were recorded. Doses measured by. the MOSkins were within 5% of the dose determined by the treatment planning system (TPS), ranging between 208 and 332 cGy, depending on dosimeter position on the rectal catheter. The overall average dose difference between measured and TPS values was 2.09% +/- 1.15% (ranging between 0.83 and 4.27%, with measured values always higher than TPS dose), subdivided in 1.40 +/- 0.37% and 2.79 +/- 1.27% for single and multichannel applicator experiments, respectively. An overall dose agreement between the TPS and measured values, detector reproducibility, and practicality of the rectal catheter demonstrated the suitability of the proposed method for in vivo real time QA purposes in gynecological HDR brachytherapy.

  • 出版日期2017-11