摘要

This paper describes a selective and ultra-sensitive analytical method for simultaneous determination of 11 fluoroquinolone (FQ) antibiotics in environmental and wastewater samples. The method employs offline solid-phase extraction (SPE) and reversed-phase high performance liquid chromatography with fluorescence detection (HPLC-FLD). A weak cation exchange SPE protocol was developed with a novel loading volume optimization algorithm and a methanol cleanup step to remove background organic matter. Various parameters were optimized to recover FQs from water/wastewater and analyte recovery was generally greater than 80%. Chromatographic separation of the 11 FQs was achieved on a 150 mm pentafluorophenyl column using a gradient elution scheme with methanol, acetonitrile, and 20 mM phosphate buffer (pH = 2.4). Excitation and emission wavelengths were individually optimized for each FQ using fluorescence spectroscopy; the excitation and emission wavelengths were 276-296 nm and 444-506 nm, respectively. Instrumental quantitation limits were 20-100 pg of mass injected. Of the 11 FQs investigated, seven (i.e., ciprofloxacin, difloxacin, enrofloxacin, fleroxacin, norfloxacin, moxifloxacin, and ofloxacin) were detected during a four-month sampling campaign of wastewater and wastewater-impacted surface water. Concentrations of FQs in raw wastewater, wastewater effluent, and wastewater-impacted surface water were 5-1292, 2-504, and 4-187 ng/L, respectively.

  • 出版日期2015-1-23