Diamagnetic measurements by concentric loops in the HL-2A tokamak

作者:Ji X Q*; Yang Q W; Xu Y; Sun T F; Yuan B S; Feng B B; Liu Y; Cui Z Y; Lu J
来源:Review of Scientific Instruments, 2013, 84(8): 083507.
DOI:10.1063/1.4816842

摘要

The diamagnetic concentric loop method in the HL-2A tokamak is described in this article. The system consists of two concentric poloidal loops with different areas enclosing the plasma column and a short time constant differential integrator, RC < 1 ms. The diamagnetic flux in HL-2A ranges from 1 mWb to 2 mWb for typical discharges with plasma current I-p = 100-400 kA. The integrator output ranges from 0.1 V to 0.2 V with time constant RC= 0.5ms, and differential area Delta S/S-out approximate to 7%. Using hybrid analog-digital compensation, the integration drift can be well compensated within 5 mV/10 s, which can meet the requirement of the concentric loop system. In this method, the measurement of differential area Delta S is not required. The vacuum toroidal flux can be compensated by adjusting the resistance in the integration circuit for several discharges with toroidal field only, which minimizes the additional error produced by a measurement of differential area. The diamagnetic concentric loop system improved the signal to noise ratio by using the short time constant integration. The system with a resolution of +/- 0.2 kJ can be used to study rapid changes in plasma stored energy, such as the additional power absorbed by the plasma, and the energy loss caused by edge localized modes.

  • 出版日期2013-8
  • 单位核工业西南物理研究院