Stability of User-Friendly Blood Typing Kits Stored Under Typical Military Field Conditions

作者:Bienek Diane R*; Chang Cheow K; Charlton David G
来源:Military Medicine, 2009, 174(10): 1075-1080.
DOI:10.7205/milmed-d-00-0909

摘要

To help preserve in-theater strength within deployed military units, commercially available, rapid, user-friendly ABO-Rh blood typing kits were evaluated to determine their stability in storage conditions commonly encountered by the warfighter. Methods for environmental exposure testing were based on MIL-STD-810F. When Eldon Home Kits 2511 were exposed to various temperature/relative humidity conditions, the results were comparable to those obtained with the control group and those obtained with industry-standard methods. For the ABO-Rh Combination Blood Typing Experiment Kits, 2 of the exposure treatments rendered them unusable. In addition, a third set of exposure treatments adversely affected the kits, resulting in similar to 30% blood type misclassifications. Collectively, this evaluation of commercial blood typing kits revealed that diagnostic performance can vary between products, lots, and environmental storage conditions.

  • 出版日期2009-10