摘要
To help preserve in-theater strength within deployed military units, commercially available, rapid, user-friendly ABO-Rh blood typing kits were evaluated to determine their stability in storage conditions commonly encountered by the warfighter. Methods for environmental exposure testing were based on MIL-STD-810F. When Eldon Home Kits 2511 were exposed to various temperature/relative humidity conditions, the results were comparable to those obtained with the control group and those obtained with industry-standard methods. For the ABO-Rh Combination Blood Typing Experiment Kits, 2 of the exposure treatments rendered them unusable. In addition, a third set of exposure treatments adversely affected the kits, resulting in similar to 30% blood type misclassifications. Collectively, this evaluation of commercial blood typing kits revealed that diagnostic performance can vary between products, lots, and environmental storage conditions.