ACOUSTIC RADIATION FORCE IMPULSE AND SUPERSONIC SHEAR IMAGING VERSUS TRANSIENT ELASTOGRAPHY FOR LIVER FIBROSIS ASSESSMENT

作者:Sporea Ioan*; Bota Simona; Jurchis Ana; Sirli Roxana; Gradinaru Tascau Oana; Popescu Alina; Ratiu Iulia; Szilaski Milana
来源:Ultrasound in Medicine and Biology, 2013, 39(11): 1933-1941.
DOI:10.1016/j.ultrasmedbio.2013.05.003

摘要

Our study compared three elastographic methods-transient elastography (TE), acoustic radiation force impulse (ARFI) imaging and supersonic shear imaging (SSI)-with respect to the feasibility of their use in liver fibrosis evaluation. We also compared the performance of ARFI imaging and SSI, with TE as the reference method. The study included 332 patients, with or without hepatopathies, in which liver stiffness was evaluated using TE, ARFI and SSI. Reliable measurements were defined as a median value of 10 (TE, ARFI imaging) or 5 (SSI) liver stiffness measurements with a success rate >= 60% and an interquartile range interval <30%. A significantly higher percentage of reliable measurements were obtained using ARFI than by using TE and SSI: 92.1% versus 72.2% (p < 0.0001) and 92.1% versus 71.3% (p < 0.0001). Higher body mass index and older age were significantly associated with inability to obtain reliable measurements of liver stiffness using TE and SSI. In 55.4% of patients, reliable liver stiffness measurements were obtained using all three elastographic methods, and ARFI imaging and TE were similarly accurate in diagnosing significant fibrosis and cirrhosis, with TE as the reference method. (E-mail: isporea@umft.

  • 出版日期2013-11