摘要

A lack of initial stability of the fixation is associated with aseptic loosening of the tibial components of cementless knee prostheses. With sufficient stability after surgery, minimal relative motion between the prosthesis and bone interfaces allows osseointegation to occur thereby providing a strong prosthesis-to-bone biological attachment. Finite element modelling was used to investigate the bone-prosthesis interface micromotion and the relative risk of aseptic loosening. It was anticipated that by prescribing different joint loads representing gait and other activities, and the consideration of varying tibial-femoral contact points during knee flexion, it would influence the computational prediction of the interface micromotion. In this study, three-dimensional finite element models were set up with applied loads representing walking and stair climbing, and the relative micromotions were predicted. These results were correlated to in-vitro measurements and to the results of prior retrieval studies. Two load conditions, (i) a generic vertical joint load of 3 x body weight with 70%/30% M/L load share and antero-posterior/medial-lateral shear forces, acted at the centres of the medial and lateral compartments of the tibial tray, and (ii) a peak vertical joint load at 25% of the stair climbing cycle with corresponding antero-posterior shear force applied at the tibial-femoral contact points of the specific knee flexion angle, were found to generate interface micromotion responses which corresponded to in-vivo observations. The study also found that different loads altered the interface micromotion predicted, so caution is needed when comparing the fixation performance of various reported cementless tibial prosthetic designs if each design was evaluated with a different loading condition.

  • 出版日期2010-4-19