A Real-Time Programmable Pulsatile Flow Pump for In Vitro Cardiovascular Experimentation

作者:Mechoor Rahul Raj; Schmidt Tyler; Kung Ethan*
来源:Journal of Biomechanical Engineering, 2016, 138(11): 111002.
DOI:10.1115/1.4034561

摘要

Benchtop in vitro experiments are valuable tools for investigating the cardiovascular system and testing medical devices. Accurate reproduction of the physiologic flow waveforms at various anatomic locations is an important component of these experimental methods. This study discusses the design, construction, and testing of a low-cost and fully programmable pulsatile flow pump capable of continuously producing unlimited cycles of physiologic waveforms. It consists of a gear pump actuated by an AC servomotor and a feedback algorithm to achieve highly accurate reproduction of flow waveforms for flow rates up to 300 ml/s across a range of loading conditions. The iterative feedback algorithm uses the flow error values in one iteration to modify the motor control waveform for the next iteration to better match the desired flow. Within four to seven iterations of feedback, the pump replicated desired physiologic flow waveforms to within 2% normalized RMS error (for flow rates above 20 mL/s) under varying downstream impedances. This pump device is significantly more affordable (similar to 10% of the cost) than current commercial options. More importantly, the pump can be controlled via common scientific software and thus easily implemented into large automation frameworks.

  • 出版日期2016-11