Determination of antimony by using tungsten trap atomic absorption spectrometry

作者:Titretir Serap; Kenduezler Erdal; Arslan Yasin; Kula Ibrahim; Bakirdere Sezgin; Ataman O Yavuz*
来源:Spectrochimica Acta Part B: Atomic Spectroscopy , 2008, 63(8): 875-879.
DOI:10.1016/j.sab.2008.03.021

摘要

An electrically heated tungsten coil was used as a trap in the determination of antimony. The technique consists of three steps. Initially, SbH(3) is formed by hydride generation procedure; then the analyte species in vapor form are transported to W-coil trap heated at 370 degrees C. Following the preconcentration step, the trap is heated to 895 degrees C; analyte species are revolatilized and transported to the flame-heated quartz atom cell where atornization and the formation of signal take place. The experimental parameters were optimized both for trap and no-trap studies. The most important experimental parameters are concentrations of HCI and NaBH4 solutions. H2 and At gas flow rates. and collection and revolatilization temperatures of W-coil. Accuracy was tested using a certified reference material, waste water EU-L-1. Limit of detection for the system is 16 ng 1- 1 using a sample of 36 ml collected in 4.0 min. Enhancement factor in sensitivity was 17.

  • 出版日期2008-8