The change in light intensity that takes place when an ambient air sample is drawn into the detection chamber of a chemiluminescence monitor generates changes in the concentrations of several species, such as NO2, NO and O-3. Although this phenomenon has been known for several decades, there is still no commonly accepted approach on when or how to correct for it in NO2 and O-3 readings.
In this work, we have assessed the expanded uncertainty of two chemiluminescence NO (x) analysers commercially available according to EN 14211:2005, with the aim of establishing the maximum allowable standard uncertainty due to the reaction between NO and O-3 in the sampling system.
Although this maximum allowable uncertainty cannot be a universal value-as it will depend on the performance of each analyser-our results have led us to propose the conservative value of 2%. We have also proposed a methodology for improving data quality which could be easily implemented by those responsible for air quality data validation.
- 出版日期2011-3