Efficient system modeling for a small animal PET scanner with tapered DOI detectors

作者:Zhang Mengxi; Zhou Jian; Yang Yongfeng; Rodriguez Villafuerte Mercedes; Qi Jinyi*
来源:Physics in Medicine and Biology, 2016, 61(2): 461-474.
DOI:10.1088/0031-9155/61/2/461

摘要

A prototype small animal positron emission tomography (PET) scanner for mouse brain imaging has been developed at UC Davis. The new scanner uses tapered detector arrays with depth of interaction (DOI) measurement. In this paper, we present an efficient system model for the tapered PET scanner using matrix factorization and a virtual scanner geometry. The factored system matrix mainly consists of two components: a sinogram blurring matrix and a geometrical matrix. The geometric matrix is based on a virtual scanner geometry. The sinogram blurring matrix is estimated by matrix factorization. We investigate the performance of different virtual scanner geometries. Both simulation study and real data experiments are performed in the fully 3D mode to study the image quality under different system models. The results indicate that the proposed matrix factorization can maintain image quality while substantially reduce the image reconstruction time and system matrix storage cost. The proposed method can be also applied to other PET scanners with DOI measurement.

  • 出版日期2016-1-21