This study aimed to evaluate the precision and accuracy of 3D reconstruction of UKA component position, contact location and lower limb alignment in standing position using biplanar radiograph. Two human specimens with 4 medial UKAs were implanted with beads for radiostereometric analysis (RSA). The specimens were frozen in standing position and CT-scanned to obtain relative positions between the beads, bones and UKA components. The specimens were then imaged using biplanar radiograph (EOS). The positions of the femur, tibia, UKA components and UKA contact locations were obtained using RSA and EOS-based techniques. Intraclass correlation coefficient (ICC) was calculated for inter-observer reliability of the EOS technique. The average (standard deviation) of the differences between two techniques in translations and rotations were less than 0.18 (0.29) mm and 0.39 degrees (0.66 degrees) for UKA components. The root-mean-square-errors (RMSE) of contact location along the anterior/posterior and medial/lateral directions were 0.84 mm and 0.30 mm. The RMSEs of the knee rotations were less than 1.70 degrees. The ICCs for the EOS-based segmental orientations between two raters were larger than 0.98. The results suggest the EOS-based 3D reconstruction technique can precisely determine component position, contact location and lower limb alignment for UKA patients in weight-bearing standing position.