New simple image overlay system using a tablet PC for pinpoint identification of the appropriate site for anastomosis in peripheral arterial reconstruction

作者:Mochizuki Yasuaki; Hosaka Akihiro*; Kamiuchi Hiroki; Nie Jun Xiao; Masamune Ken; Hoshina Katsuyuki; Miyata Tetsuro; Watanabe Toshiaki
来源:Surgery Today, 2016, 46(12): 1387-1393.
DOI:10.1007/s00595-016-1326-4

摘要

To evaluate the accuracy and utility of a new image overlay system using a tablet PC for patients undergoing peripheral arterial reconstruction. Eleven limbs treated with distal bypass surgery were studied. Three-dimensional images obtained by processing a preoperative contrast-enhanced computed tomography scan were superimposed onto the back-camera images of a tablet PC. We used this system to pinpoint a planned distal anastomotic site preoperatively and to make a precise incision directly above it during surgery. We used a branch artery near the distal anastomotic site as a reference point and the accuracy of the system was validated by comparing its results with the intraoperative findings. The precision of the system was also compared with that of a preoperative ultrasonographic examination. Both the image overlay system and ultrasonography (US) accurately identified the target branch artery in all except one limb. In that limb, which had a very small reference branch artery, preoperative US wrongly identified another branch, whereas the image overlay system located the target branch with an error of 10 mm. Our image overlay system was easy to use and allowed us to precisely identify a target artery preoperatively. Therefore, this system could be helpful for pinpointing the most accurate incision site during surgery.

  • 出版日期2016-12