Ultrasonic Elastography in Breast Cancer Diagnosis: Strain Ratio vs 5-point Scale

作者:Zhi, Hui*; Xiao, Xiao-Yun; Yang, Hai-Yun; Ou, Bing; Wen, Yan-Ling; Luo, Bao-Ming
来源:Academic Radiology, 2010, 17(10): 1227-1233.
DOI:10.1016/j.acra.2010.05.004

摘要

Rationale and Objectives: The aim of this study was to develop a more reliable ultrasonic elastographic diagnostic method than a five-point scoring system by analyzing the difference in stiffness between benign and malignant breast lesions. @@@ Materials and Methods: From January 2008 to April 2009, 559 solid lesions (415 benign, 144 malignant) in 437 consecutive patients (age range, 12-77 years) were examined using ultrasound elastography (UE). Final diagnosis was made on the basis of histopathologic findings. The strain ratios of the lesions were calculated. The area under the curve and cutoff point, both of which were obtained using receiver-operating characteristic curve analysis, were used to assess diagnostic performance. Diagnostic performance was further compared to that generated using a five-point scoring system with the z test. The sensitivity, specificity, and accuracy of these two evaluation systems were compared using McNemar's test. @@@ Results: The strain ratios of benign lesions (mean, 1.83 +/- 1.22) and malignant lesions (mean, 8.38 +/- 7.65) were significantly different (P < .00001). When a cutoff point of 3.05 was introduced, UE had 92.4% sensitivity, 91.1% specificity, and 91.4% accuracy. The area under the curve for strain ratio-based elastographic analysis was 0.944, and the area under the curve for the five-point scoring system was 0.885. The diagnostic performance of strain ratio-based elastographic analysis was better than that of the five-point scoring system with UE (P < .05). @@@ Conclusions: Strain ratio-based elastographic analysis can provide a new, more reliable diagnostic tool in comparison to a five-point scoring system for UE.