摘要

Background: Reference laboratories providing reference services recognized by the Joint Committee for Traceability in Laboratory Medicine (JCTLM) must be accredited as calibration laboratories according to ISO 17025 and ISO 15195. These standards require laboratories to establish an uncertainty budget, in which the uncertainty contributions of the relevant uncertainty components are specified. We present a model to estimate the measurement uncertainty of creatine kinase catalytic activity concentration results obtained by IFCC primary reference measurement procedure. Methods: The measurement uncertainty has been estimated by following the next steps: 1) specification of the measurand; 2) identification of the most relevant uncertainty sources; 3) estimation of standard uncertainties by either type A or type B evaluation; 4) estimation of combined uncertainty while taking into account sensitivity coefficients, as well as existence of correlated uncertainty sources; and 5) estimation of expanded uncertainty with a defined coverage probability. Results: The estimated expanded uncertainty was 2.2% (k = 2). Uncertainty sources with a significant contribution to the measurement uncertainty were the following: pH adjustment (0.68%), absorbance accuracy (0.48%), wavelength adjustment (0.20%), reaction temperature (0.19%), volume fraction of sample (0.15%) and absorbance linearity (0.06%). Conclusions: The present model is an approach to establish the uncertainty budget of primary reference procedures for the measurement of the catalytic activity concentration of enzymes, and aims at being an example to be followed by other reference laboratories, as well as by laboratories that carry out primary reference measurement procedures.

  • 出版日期2015-4