Computational Study of Aqueous Humor Dynamics Assessing the Vault and the Pupil Diameter in Two Posterior-Chamber Phakic Lenses

作者:Ignacio Fernandez Vigo Jose; Macarro Merino Ana; Fernandez Francos Joaquin; De Pablo Gomez de Liano Lucia; Maria Martinez de la Casa Jose; Garcia Feijoo Julian; Angel Fernandez Vigo Jose
来源:Investigative Ophthalmology & Visual Science, 2016, 57(11): 4625-4631.
DOI:10.1167/iovs.16-19900

摘要

PURPOSE. To compare the behavior of aqueous humor (AH) and analyze flow differences by comparing the volume and velocity of the flow after two different models of implantable collamer lens (ICL) placement. METHODS. Computational fluid dynamics with numerical simulation using Ansys Fluent software was performed to compare the AH flow through a peripheral iridotomy (PI), which is typically performed after implantation of a V4b lens to the central hole of a V4c lens. The volume and flow rate in 24 scenarios were compared according to the type of lens, pupil diameter (PD) (3.5 or 5.5 mm), the vault (100, 350, and 800 mu m) and the PI (single or double, 180 or 360 mu m). RESULTS. With a standard vault (350 mu m) and a PD of 3.5 mm, the volume of AH that flows from the posterior to the anterior chamber through the PI (V4b lens: 73.4% in 360 mu m and 17.3% in 180 mu m) and for the central hole (V4c lens: 75.7%) is larger than in the case of a PD of 5.5 mm (13.9%, 0.91%, and 15.3% respectively). When the vault is low (100; PD 3.5 mm), the volume of AH that reaches the central hole of the V4c lens is diminished (52.0%), being 5.1% if the pupil is enlarged. CONCLUSIONS. AH flow varies depending on the type of ICL implanted, whether it is implanted with an iridotomy or a central hole on the lens, the PD, and the vault.

  • 出版日期2016-9