The objective of the present work is the application of the Monte Carlo method (GUMS1) for evaluating uncertainty in electromagnetic field measurements and the comparison of the results with the ones obtained using the 'standard' method (GUM). In particular, the two methods are applied in order to evaluate the field measurement uncertainty using a frequency selective radiation meter and the Total Exposure Quotient (TEQ) uncertainty. Comparative results are presented in order to highlight cases where GUMS1 results deviate significantly from the ones obtained using GUM, such as the presence of a non-linear mathematical model connecting the inputs with the output quantity (case of the TEQ model) or the presence of a dominant nonnormal distribution of an input quantity (case of U-shaped mismatch uncertainty). The deviation of the results obtained from the two methods can even lead to different decisions regarding the conformance with the exposure reference levels.