Comparison of invasive and oscillometric blood pressure measurement techniques in anesthetized camelids

作者:Aarnes Turi K*; Hubbell John A E; Lerche Phillip; Bednarski Richard M
来源:Canadian Veterinary Journal, 2012, 53(8): 881-885.

摘要

This study assessed the accuracy of the oscillometric method for arterial blood pressure (ABP) monitoring in anesthetized camelids. Twenty camelids were anesthetized and systolic ABP (SABP), mean ABP (MABP), and diastolic ABP (DABP) were measured directly and using the oscillometric method. The mean difference between SABP measurements was -9.9 +/- 21.9 mmHg with a range of -76 to 54 mmHg, and the 95% limits of agreement (LOA) were -33 to 53 mmHg. The difference between DABP measurements was -1.8 +/- 15.6 mmHg with a range of -81 to 36 mmHg, and the 95% LOA were -32 to 29 mmHg. The difference between MABP measurements was -2.9 +/- 17.0 mmHg with a range of -81 to 36 mmHg, and the 95% LOA were -30 to 36 mmHg. Accurate ABP monitoring in anesthetized camelids cannot be accomplished using the oscillometric method.

  • 出版日期2012-8