A case-control study evaluating RT-PCR/ESI-MS technology compared to direct fluorescent antibody and xTAG RVP PCR

作者:Hardick Justin*; Sadiq Sufyan; Perelstein Elizabeth; Peterson Stephen; Rothman Richard; Gaydos Charlotte A
来源:Diagnostic Microbiology and Infectious Disease, 2014, 79(2): 187-189.
DOI:10.1016/j.diagmicrobio.2014.02.009

摘要

Waste nasopharyngeal swabs (N = 244) were evaluated by the reverse-transcriptase polymerase chain reaction/electrospray ionization mass spectrometry PLEX-ID Broad Respiratory Virus Surveillance Kit version 2.5 compared to direct fluorescent antibody and xTAG Respiratory Virus Panel for percent agreement, sensitivity, and specificity. Sensitivity and specificity were 91% (111/122) and 95.1% (116/122), respectively. Sensitivity by virus, except parainfluenza, was 92.9-100%, and specificity was 99-100%.

  • 出版日期2014-6