摘要

Aim: To resolve the event in tissue Doppler (TDI)- and speckle tracking-based velocity/time curves that most accurately represent aortic valve closure (AVC) in infarcted ventricles and at high heart rates. Methods: We studied the timing of AVC in 13 patients with myocardial infarction and in 8 patients at peak dobutamine stress echo. An acquisition setup for recording alternating B-mode and TDI image frames was used to achieve the same frame rate in both cases (mean 136.7 frames per second [FPS] for infarcted ventricles, mean 136.9 FPS for high heart rates). The reference method was visual assessment of AVC in the high frame rate narrow sector B-mode images of the aortic valve. Results: The initial negative velocities after ejection in the velocity/time curves occurred before AVC, 44.9 +/- 21.0 msec before the reference in the high heart rate material, and 25.2 +/- 15.2 msec before the reference in the infarction material. Using this time point as a marker for AVC may cause inaccuracies when estimating end-systolic strain. A more accurate but still a practical marker for AVC was the time point of zero crossing after the initial negative velocities after ejection, 5.4 +/- 15.3 msec before the reference in high heart rates and 8.2 +/- 12.9 msec after the reference in the infarction material. Conclusion: The suggested marker of AVC at high heart rate and in infarcted ventricles was the time point of zero crossing after the initial negative velocities after ejection in velocity/time curves. (Echocardiography 2010;27:363-369).