摘要

In this paper, we have designed a signal amplified method for the electrochemical determination of polynucleotide kinase activity. It is based on (a) the peroxidase-like activity of magnetite microspheres (MNPs), (b) the specific recognition capabilities of titanium dioxide (TiO2) with the phosphate groups of the capture probe and (c) the DNA dendrimer structure for signal amplification. MNPs coated with TiO2 (TMNPs) were prepared and characterized by scanning electron microscopy (SEM) and Fourier transform infrared (FTIR) spectroscopy. TMNP-DNA dendrimers were formed by the hybridization of captured nucleic acids with a link probe. Cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were carried out to study the electrocatalytic process. The formation of the TMNP-DNA dendrimer structures was related to the phosphorylated capture probe and further to the activity of polynucleotide kinase, which was the base of the polynucleotide kinase detection. The TMNP-DNA dendrimer based biosensor showed sensitive detection of polynucleotide kinase with a satisfying result; a low detection of 0.003 U mL(-1) and wide linear range of 0.01 to 30 U mL(-1) were achieved. Additionally, the present TMNP-DNA dendrimer based biosensor also demonstrated excellent selectivity, stability and reproducibility.