A Wearable Thermometry for Core Body Temperature Measurement and Its Experimental Verification

作者:Huang Ming*; Tamura Toshiyo; Tang Zunyi; Chen Wenxi; Kanaya Shigehiko
来源:IEEE Journal of Biomedical and Health Informatics, 2017, 21(3): 708-714.
DOI:10.1109/JBHI.2016.2532933

摘要

A wearable thermometry for core body temperature (CBT) measurement has both healthcare and clinical applications. On the basis of the mechanism of bioheat transfer, we earlier designed and improved a wearable thermometry using the dual-heat-flux method for CBT measurement. In this study, this thermometry is examined experimentally. We studied a fast-changing CBT measurement (FCCM, 55 min, 12 subjects) inside a thermostatic chamber and performed long-term monitoring of CBT (LTM, 24 h, six subjects). When compared with a reference, the Core Temp CM-210 by Terumo, FCCM shows 0.07 degrees C average difference and a 95% CI of [-0.27, 0.12] degrees C. LTM shows no significant difference in parameters for the inference of circadian rhythm. The FCCM and LTM both simulated scenarios in which this thermometry could be used for intensive monitoring and daily healthcare, respectively. The results suggest that because of its convenient design, this thermometry may be an ideal choice for conventional CBT measurements.

  • 出版日期2017-5