A Peltier cooling diamond anvil cell for low-temperature Raman spectroscopic measurements

作者:Noguchi Naoki*; Okuchi Takuo
来源:Review of Scientific Instruments, 2016, 87(12): 125107.
DOI:10.1063/1.4972254

摘要

A new cooling system using Peltier modules is presented for a low-temperature diamond anvil cell instrument. This cooling system has many advantages: it is vibration-free, low-cost, and compact. It consists of double-stacked Peltier modules and heat sinks, where a cooled ethylene glycol-water mixture flows through a chiller. Current is applied to the Peltier modules by two programmable DC power supplies. Sample temperature can be controlled within the range 210-300 K with a precision of +/- 0.1 K via a Proportional-Integral-Differential (PID) control loop. A Raman spectroscopic study for the H2O ice VII-VIII transition is shown as an example of an application of the Peltier cooling diamond anvil cell system. Published by AIP Publishing.

  • 出版日期2016-12