A novel LC-MS approach for the detection of metabolites in DMPK studies

作者:Plumb Robert S*; Mather Joanne; Little David; Rainville Paul D; Twohig Marian; Harland Gary; Kenny Daniel J; Nicholson Jeremy K; Wilsons Ian D; Kass Ignatius J
来源:Bioanalysis, 2010, 2(10): 1767-1778.
DOI:10.4155/BIO.10.115

摘要

Background: The profiling and quantification of drug metabolites in discovery and development bioanalysis studies is playing an increasingly important role in early candidate selection. Using a conventional tandem quadrupole mass spectrometer this activity normally requires several analytical runs to acquire the necessary analytical data. Results: In this article we present the use of a new tandem quadrupole mass spectrometer equipped with a novel collision cell design, which allows the rapid switching between multiple reaction monitoring and full-scan MS mode. This approach allowed for the collection of multiple reaction monitoring data and full-scan data with no loss in sensitivity, with analysis times in the 1-2 min range. Conclusion: A modified approach of using the multiple reaction monitoring data to trigger the acquisition of full scan MS/MS data is described, where the data is collected on the trailing edge of the LC-MS peak, thus improving data quality and throughput.

  • 出版日期2010-10