摘要

Measurement of amide H/D exchange on the ms time scale can provide valuable information about the dynamic behavior of the most flexible regions of proteins. We describe here a simple mixing apparatus, assembled solely from off-the-shelf components, that can be used for HID exchange mass spectrometry to measure exchange on the 50-5000 ms time scale. Our apparatus utilizes flow-injection to minimize sample consumption. Although the mixer operates at low Reynolds numbers (less than 10(2)) where laminar flow is expected, H/D exchange kinetics were well-approximated using the assumption of plug-flow. We validated this approximation using fluorescence imaging of fluorescein-conjugated bovine serum albumin in the delay line and by demonstrating agreement between measured and calculated H/D exchange kinetics for a mixture of peptides. The performance of the apparatus was further validated by measuring rapid HID exchange kinetics by an intrinsically disordered protein, murine CBP2059-2117 (UniProt CBP_MOUSE). H/D exchange data from CBP, both free and in complex with human ACTR(1018-1688) (UniProt NCOA3_HUMAN), were consistent with previous biophysical studies of this protein.

  • 出版日期2013-5-21