摘要

An octyl-functionalized hybrid silica monolithic column was developed for in-tube solid-phase microextraction (SPME) to perform on-line pre-concentration coupled to capillary high-performance liquid chromatography (mu HPLC) analysis. A hybrid silica monolithic column functionalized with octyl groups was conveniently synthesized by a two-step acidibase-catalyzed hydrolysis/co-condensation of tetraethoxysilane (TEOS) and n-octyltriethoxysilane (C8-TEOS). The size of through-pores as well as the carbon content can be adjusted by changing the ratio of TEOS to C8-TEOS in the polymerization mixture. The extraction characteristics of the monolithic column prepared under optimized fabrication conditions were studied by using polycyclic aromatic hydrocarbons (PAHs) as the analytes. The sample volume that could be injected into the system was increased up to 1 mL with simultaneous increase of column efficiency, when hybrid silica monolithic column was used as a precolumn. Good linear calibration curves (R > 0.999) were obtained, and the limits of detection (signal-to-noise ratio, S/N = 3) for the analytes were found to be between 2.4 and 8.1 ng/mL with a UV absorbance detector, which are 299-456 times lower than those obtained without preconcentration. The column-to-column RSD values were 1.3-8.0% for recoveries of PAHs investigated.